- Packing Material Differences - Hypersil
- Packing Material Selection - Inertsil
- Fastener Seperations: Resolution
vs. Column Length
- Kromasil
- Sensitivity/Solvent Sonsumption: 3mm vs. 4.6mmID - Monitor
- Column Selection Criteria - Nucleosil
- pH Protection: Why use Gaurd Cartridges? - Resiasil
- USP Categories For HPLC Columns - Spherisorb

Faster Separations: Resolution vs. Column Length

5cm.

10cm.

Sample: uracil, acetophenone, methyl benzoate, toluene, naphthalene

15cm.

25cm.

Sensitivity/Solvent Consumption: 3mmID vs. 4.6mmID

 

System:

Beckman Altex 110 Pump

  Linear UV-106 Detector
  Tubing: 0.007"ID Inlet - 0.005"ID Outlet
  Detection: UV @ 254
  Setting: 0.2 AUFS
Back Pressure: about 1200psi for both columns
Mobile Phase: 60% Acetonitrile
40% Water
Sample Volume: 5uL
Sample: Uracil Phenol "N,N-Dimethylaniline" Naphthalene
Material: Inertsil 5um ODS2

4.6mmID

3mmID

Different Variables For Each Chromatogram

Flow Rate: 1.0mL/min.(4.6mmID)
0.5mL/min.(3mmID)
Linear Velocity As Follows:
6.02cm/min. (4.6mmID)
7.07cm/min. (3mmID)
Sensitivity: See below; setting @ 0.2 AUFS
Solvent Usage: 60mL./hr. (4.6mmID) 30mL/hr. (3mmID)
Column Dimensions: 4.6mmID (150x4.6mmID) 3mmID (150x3mmID)
   
   

 

 